Cetane Number Analyser In UAE

Cetane Number Analyzer

Cetane Number Analyzer measures diesel fuel ignition delay operating per ASTM D613. It incorporates constant-volume combustion chambers, piezoelectric pressure transducers, dynamic fuel injection controls, and cetane rating software.

Warranty

Installation

Training

Product overview

The AFIDA Cetane Number Analyser (ASTM D613) is a revolutionary development providing you with a fully automated way of determining Indicated Cetane Number (ICN) of diesel and diesel-related fuels.

The AFIDA Cetane Number Analyser (ASTM D613) is a revolutionary development providing you with a fully automated way of determining Indicated Cetane Number (ICN) of diesel and diesel-related fuels. The AFIDA cetane analyser can analyse multiple samples in sequence and selects the samples using the 36-place autosampler. Only a small sample volume of 40 ml is required per test, with a short analysis time of approximately 25 minutes.

The analyser uses a Constant Volume Combustion Chamber (CVCC) and incorporates a unique and patented high-pressure injection system that generates fine fuel droplets similar to modern common rail injectors in most diesel engines. making it a reliable cetane number analyzer aligned with ASTM D613 testing requirements.

An ultra-robust temperature-controlled piezo-electric injector provides rapid switching and highly repeatable fuel metering, offering you improved performance and consistency when compared with solenoid-controlled injectors and pintle-type nozzles.

As a leading Cetane Number Analyzer supplier, we are dedicated to providing advanced fuel testing infrastructure and end-to-end technical support across the Middle East, serving key markets including the UAE, Saudi Arabia, Qatar, Kuwait, Oman, and Bahrain.

 

ASTM D1655 Standard Specification for Aviation Turbine Fuels
ASTM D975 Standard Specification for Diesel Fuel Oils , biodiesel, biodiesel blend, diesel, fuel oil, petroleum and petroleum products
IP 617 Determination of generic cetane number (G-CN) of fuels using a constant volume combustion chamber: primary reference fuels calibration (PRFC) method
BS EN 17155 Liquid petroleum products – determination of indicated cetane number (ICN) of middle distillate fuels – primary reference fuels calibration method using a constant volume combustion chamber
ASTM D613 Standard Test Method for Cetane Number of Diesel Fuel Oil
IP 41 Determination of the ignition quality of diesel fuels — Centane engine method
BS EN ISO 5165 & BS 2000-41 & ISO 5165 Determination of the ignition quality of diesel fuels. Cetane engine method
ASTM D8183 Determination of Indicated Cetane Number (ICN) of Diesel fuel oils using a Constant Volume Combustion Chamber – reference fuels calibration method
ASTM D6751 Standard Specification for Biodiesel Fuel Blend Stock (B100) for Middle Distillate Fuels
ASTM D7467 Standard Specification for Diesel Fuel Oil, Biodiesel Blend (B6 to B20)
EN 590 Automotive fuels – Diesel – Requirements and test methods
EN 15940 Automotive fuels – Paraffinic diesel fuel from synthesis or hydrotreatment – Requirements and test methods
ASTM D4054 Standard Practice for Evaluation of New Aviation Turbine Fuels and Fuel Additives

 

As a state-of-the-art Cetane Analyzer, the instrument boasts advanced operational benefits:

  • Excellent correlation to ASTM D613, IP 41 and BS EN ISO 5165, BS 2000-41, ISO 5165, ensuring it acts as a dependable ASTM D613 Cetane Number tester.
  • Wide sample types – middle distillates, blend components, biodiesel, FAME, XTL, HVO, heating oil and Jet A1
  • Suitable for use with Cetane improvers
  • Calibrated with industry-approved PRFs over the range 35 to 85 ICN
  • Precise and reproducible software-controlled determination of ICN
  • Results saved to PC or for download
  • Stable calibration – fixed test conditions eliminate routine temperature tuning
  • Low running costs, single gas supply for combustion air

 

Chamber temperature 580 °C
Chamber pressure 17.5 bar
Injection pressure 1000 bar
Compressed air 20.9 ± 0.5 % O2
Sample volume approx. 40 ml for analysis and cleaning
Warm up time approx. 45 min
Analysis time approx. 25 min per sample
Size (DxHxW) 130 cm x 80 cm x 60 cm (excluding platform)
Weight 160 kg 100 kg (excluding platform)
CCCN Code Tariff 90318080
Temperature range 0 to 120 °C (no external cooling required). Test the sample for multiple temperatures (ASTM D6377)
Pressure range 0-1000 kPa
Pressure resolution 0.001 kPa
Units of pressure kPa, psi
Temperature accuracy better than 0.05 °C
Vapour to liquid ratio 0.02:1 to 20:1. Test the sample for multiple V/L ratios (ASTM D6377)
Sample volume 1 ml, plus rinsing
Measurement time 5 minutes for D5191
Lubrication Automatic
Voltage 100/240 Vac, 50/60 Hz, Auto-sensing univeral power supply

 

  • Petrochemical
  • Research
  • Refineries
  • testing laboratories