SEAL’s BD50 and BD28 laboratory digestion systems automate in high-temperature (over 380°C).
The SEAL BD50 and BD28 laboratory digestion systems automate sample digestion to deliver the productivity, quality, and safe operation needed for a strong analytical program. These programmable Laboratory digesters perform high-temperature digestion, above 380°C, and are available in two formats: a 50-place system for 75 ml or 100 ml digestion tubes and a 28-place system for 250 ml tubes.
Built for durability and robustness, each system includes a block digestion unit, programmable controller, tube rack, draft shield, and digestion tubes, with an optional cooling stand to support the tube rack above the block. This digestion system is ideal for Kjeldahl digestion ahead of Total Kjeldahl Nitrogen (TKN) and Phosphorus (TKP) analysis and is supplied by BDH Middle East to laboratories across the United Arab Emirates (UAE), Oman, Qatar, Kuwait, Bahrain, and Saudi Arabia as a specialist supplier.
| EPA Method Number | EPA Method Name |
| 351.2 | Determination of Total Kjeldahl Nitrogen by Semi-Automated Colorimetry |
| 365.1 | Determination of Phosphorus by Semi-Automated Colorimetry |
| 365.4 | Phosphorous, Total (Colorimetric, Automated, Block Digester AA II) |
Automation
- Automatic operation frees key laboratory personnel from tedious manual digestions
- Up to 20 programs with 30 steps each can be stored, automating complex procedures
- Programmable standby temperature enables rapid start-up, increasing daily batch processing
- Overnight digestion capability improves throughput
Efficiency
- Controlled heating-rate changes as low as 1°C per minute prevent bumping and spattering of reactive samples
- No ongoing consumable costs beyond power and reagents
- Compact design saves fume hood space, with a separate controller that can operate outside the fume hood for safety and convenience
Compliance
- A unique heating element ensures even heat distribution across the solid aluminum block
- Digital storage of digestion methods ensures reproducibility between batches and operators
- Automatic power-failure sensing and recording of digest conditions protect results
- A built-in temperature cutoff system prevents overheating