Global Analyser Solutions’ PIONA GC-MS Analyzer provides an all-in-one Paraffins, Isoparaffins, Olefins, Naphthenes, Aromatics and oxygenates analysis for gasoline range fuels within 35 minutes.
-MS based methods are now widely used in petrochemical laboratories, from ASTM D5769 (aromatics in gasoline) to IP 585/10 (FAME in aviation fuel). The PIONA GC-MS Analyzer from Global Analyser Solutions offers complete analysis of paraffins, iso-paraffins, olefins, naphthenes, aromatics and oxygenates in gasoline-range materials, characterising spark-ignition fuels and similar samples by individual carbon number up to C12.
Compared with conventional multidimensional PIONA analysis, the GC-MS method needs only 35 minutes per run and reports individual components directly. A single high-resolution capillary column keeps uptime high and maintenance low, while results show excellent agreement with multidimensional systems. The analyser also extends to related applications such as FAME in Jet Fuel A1, and is available to laboratories across the United Arab Emirates (UAE), Oman, Qatar, Kuwait, Bahrain and Saudi Arabia through Global Analyser Solutions, a specialist Supplier
ASTM D6839
- Single-column analyser delivering full PIONA analysis
- Just 35 minutes analysis time
- Stable, high-uptime performance with low maintenance
- Lower running costs than multidimensional systems
- Simple method setup with individual component reporting
- Handles high olefin content reliably
- Results closely match complex multidimensional PIONA analysis
- Adaptable to other GC-MS applications, such as FAME in Jet Fuel A1
- Application: full PIONA analysis of hydrocarbon samples such as naphtha and spark-ignition fuels
- Configuration: single-channel instrument with single-column separation and MS detection
- Injector: split-splitless injector
- Column: 100m × 0.25mm ID, 0.5µ apolar phase, or 40m × 0.18mm ID, 0.5µ apolar phase
- Detection: Thermo iSQ quadrupole mass spectrometer
- Software: dedicated PIONA software with a database of 1,100 MS spectra and 400 component retention indices; automated peak identification, quantitation, reporting and deconvolution
- Analysis time: 100 minutes (100m column) or 35 minutes (40m column)