Home / Products / Testing & Measurement / Analytical Solutions / Thermogravimetric Analyzer (TGA)
Thermogravimetric Analyzer (TGA) Suppliers in UAE

Thermogravimetric Analyzer (TGA)

A thermogravimetric analyzer measures how a sample weight varies as the temperature rises and determines the exact moment when the material starts to deteriorate. A TGA analyzer uses a highly sensitive microbalance together with a programmable furnace and a purge gas control system.

Warranty

Installation

Training

Product overview

A thermogravimetric analyzer measures how a sample weight varies as the temperature rises and determines the exact moment when the material starts to deteriorate

The AMI TGA analyzer keeps weighing a sample as the furnace climbs through a set temperature program, so every mass change in the run lines up with an exact temperature and time. It comes down to three parts working together, a high resolution microbalance, a programmable furnace and a purge gas line. Swap the gas and the curve changes too.

 Run a sample under nitrogen and it shows plain thermal breakdown. Switch to air or oxygen and an oxidative step shows up that inert gas never reveals. Argon, reducing gas, pressurised gas, all of these work, and that range is exactly why one TGA thermogravimetric analyzer ends up covering so many different test methods on a single bench.

When properly interpreted, each mass loss step reveals its own data, segregating vapors, binder, polymeric backbone, carbon black and final ash in a single run. The remaining ash gives the amount of mineral filler present in the product. 

The temperature at which the process begins is termed the onset temperature, which is the temperature ascertained by the formulators. This number is of great interest to formulators. All manufacturers of various materials, from polymers companies to paint producers, apply TGA analysis, but for various materials.

Three furnace ranges

Pick ambient to 1000°C, 1200°C or 1500°C depending on whether the work is plastics and food or refractory and ceramic material.

Insulated balance housing

The balance sits behind thermal insulation, so furnace heat does not push the zero point around during a long ramp. Drift stays low and the mass reading stays honest.

Built for small changes

Resolution of 0.1 μg with isothermal drift held at or under 5 μg per hour means low level additives and trace moisture still register.

Specification RT-1000°C RT-1200°C RT-1500°C
Temperature Range RT-1000°C RT-1200°C RT-1500°C
Temperature Accuracy ±0.5°C
Temperature Precision ±0.1°C
Program Rate 0.1-300°C/min 0.1-60°C/min
Cooling Mode Water Cooling
Resolution 0.1 μg
Measuring Range ±200 mg
Dynamic Baseline Drift ≤ 25 μg (No blank background subtraction)
Isothermal Baseline Drift ≤ 5 μg/h
Repeatability ≤ 10 μg

Find the Right Solution for Your Laboratory

Need help selecting the right product for your application? Our specialists can guide you with product specifications, pricing, installation, and technical support across the UAE and Middle East.